Binder VD115 Vacuum drying oven for non-flammable solvents
A BINDER VD series chamber is impressive while gently drying with its homogeneous temperature distribution. The patented expansion shelf technology ensures optimal heat transfer. The shelves are easy to position where required and easy to dry in the inner chamber.
- Electronically controlled APT.line™ preheating chamber with 2 expansion racks assuring temperature accuracy and reproducible results
- Temperature range from 15 °C above ambient temperature to 200 °C
- MP controller with 2 programs with 10 sections each or switchable to 1 program with 20 sections
- Integrated weekly program timer with real-time function
- Digital temperature setting with an accuracy of one degree
- Elapsed time indicator - Precision-adjustable ventilation valve
- Precision-adjustable inert gas valve with Cross-Flow-Technology
- All electrical components are decoupled from the inner chamber
- Spring-mounted safety glass panel with shatter protection
- Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm
- Measuring port DN 16 in rear panel
- Analog pressure gauge (displays pressure difference between the inner chamber and ambient pressure)
- Electro polished inner chamber, suction and ventilation tubes, pressure container, expansion racks, and ball valve are made of stainless steel
- Door gasket made of tempered silicone
- 2 x 24 V DC (max 0.4 A) switching outputs, switched via 2 control contacts in the program editor
- RS 422 interface for communication software APT-COM™ DataControlSystem
- 2 patented, flexible aluminum expansion racks
- Also available as complete system with module and vacuum pump. Features:
- Reduced noise level
- Practical working height
- Coordinated system - BINDER test certificate
Technical Specifications
Exterior dimensions |
|
Width (mm) |
740 |
Height (incl. feet) (mm) |
900 |
Height with optional vacuum module (mm) |
622 |
Height with optional vacuum module (mm) |
1522 |
Depth (mm) |
670 |
Plus door handle, connection (mm) |
100 |
Wall clearance, rear (mm) |
100 |
Wall clearance, side (mm) |
135 |
Interior dimensions |
|
Width (mm) |
506 |
Height (mm) |
506 |
Depth (mm) |
450 |
Interior volume (l) |
115 |
Expansion racks (aluminum) (number standard/max.) |
2/6 |
Distance between the racks (mm) |
68 |
Usable space per rack (width x depth) (mm) |
455 x 440 |
Load per shelf (kg) |
20 |
Permitted total load (kg) |
65 |
Weight (empty) (kg) |
153 |
Temperature range |
|
Temperature range approx. 15 °C above ambient temperature to (°C) |
200 |
Temperature variation 1) |
|
100 °C (± K) |
3,5 |
200 °C (± K) |
9 |
Temperature fluctuation (± K) 1) |
0,1 |
Heating up time 1), 2) switch position II |
|
to 100 °C (min.) |
90 |
to 200 °C (min.) |
190 |
Vacuum connection with small flange (DN mm) |
16 |
Measuring access port with small flange (DN mm) |
16 |
Inert gas connection with flow limiter thread (RP) |
3/8 |
Permitted end vacuum (mbar) |
0,01 |
Leak rate (max. bar 1/h) |
0,01 |
Electrical data |
|
IP protection class acc. to EN 60529 |
IP 20 |
Voltage (± 10%) 50 / 60 Hz |
230 |
Nominal power (kW) |
1,7 |
Energy consumption |
|
at 100 °C (W) |
250 |
at 200 °C (W) |
785 |
1) Value with aluminum racks
2) To 98% of the set value
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. These average values have been determined according to the BINDER factory standard, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. Differing ambient temperatures and production-related device-specific variances can lead to varying technical data.
Therefore, we recommend individual customer-specific calibration or validation for applications on the limit of the permitted ambient temperature range.
Contact us on 01939 291200 or by email info@sciquip.co.uk