High-Temperature Ovens, Chamber Furnaces with Air Circulation


N 120/65 HA


N 250/65HA with pneumatic lift-door as additional equipment

 

 

 


Over-temperature limit controller

N 15/65HA, N 30/45HA - N 500/85HA

These chamber furnaces with air circulation are characterized by their extremely high temperature uniformity. Hence, they are especially suitable for processes such as cooling, crystalizing, pre-heating, curing, but also for numerous processes in tool making. Due to the modular concept, the furnaces can be adjusted to the process requirements by adding suitable equipment.

  • Tmax 450 °C, 650 °C, or 850 °C
  • Horizontal air circulation
  • Swing door hinged on the right
  • Optimum temperature uniformity (according to DIN 17052-1) up to ?T 6 K
  • Heating from bottom, sides and top
  • Optimum air flow and temperature uniformity through high circulation rates
  • One shelf and rails for two additional shelves included
  • Air baffle box of stainless steel inside the furnace chamber for optmum air circulation
  • Switchgear with solid-state relays

 

Additional equipment N 30/45HA - N 500/85HA

  • Cooling system with fan for fast cooling
  • Automatic control of air inlet and exhaust air vents for better ventilation and faster cooling of the furnace
  • Pneumatic lift-door
  • Atmosphere boxes for protective gas operation
  • Additional shelves
  • Variable frequency control of air circulation velocity for sensitive parts
  • Charging aids, roller conveyors

 

Technical Specifications
Model
Tmax°C
Inner dimensions (mm)
Volume
in L
Outer dimensions (mm)
Supply
Power kW
Electrical
Connections
Weight
(kg)
w
d
h
w
d
h
N 30/45 HA
450
290
420
260
30
607 + 255
1175
1315
3,6
1-phase
195
N 60/45 HA
450
350
500
350
60
667 + 255
1250
1400
6,6
3-phase
240
N 120/45 HA
450
450
600
450
120
767 + 255
1350
1500
9,6
3-phase
310
N 250/45 HA
450
600
750
600
250
1002 + 255
1636
1860
19,0
3-phase
610
N 500/45 HA
450
750
1000
750
500
1152 + 255
1886
2010
28,0
3-phase
1030
N 15/65 HAš
650
295
340
170
15
470
845
460
2,7
1-phase
55
N 30/65 HA
650
290
420
260
30
607 + 255
1175
1315
6,0
3-phase2
195
N 60/65 HA
650
350
500
350
60
667 + 255
1250
1400
9,6
3-phase
240
N 120/65 HA
650
450
600
450
120
767 + 255
1350
1500
13,6
3-phase
310
N 250/65 HA
650
600
750
600
250
1002 + 255
1636
1860
21,0
3-phase
610
N 500/65 HA
650
750
1000
750
500
1152 + 255
1886
2010
31,0
3-phase
1030
N 30/85 HA
850
290
420
260
30
607 + 255
1175
1315
6,0
3-phase2
195
N 60/85 HA
850
350
500
350
60
667 + 255
1250
1400
9,6
3-phase
240
N 120/85 HA
850
450
600
450
120
767 + 255
1350
1500
13,6
3-phase
310
N 250/85 HA
850
600
750
600
250
1002 + 255
1636
1860
21,0
3-phase
610
N 500/85 HA
850
750
1000
750
500
1152 + 255
1886
2010
31,0
3-phase
1030

 

šTable-top model ˛Heating only between two phases

 

 

 