FDL 115
Safety drying oven with first class safety features
This series presents the perfect environment for solvent-based samples. The interior chamber has a symmetrical airflow with defined flow velocities, free of silicone and dust, safeguarded by a high-efficiency filter cartridge. FDL thus complies with all EN 1539 and ISO 3215 requirements and its program with the intelligent fresh air monitoring feature offers maximum occupational safety.
- Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
- All safety features acording to EN 1539
- Temperature range 5 °C (32 ºF) above ambient temperature up to 300 °C (572 °F)
- MP Controller, 2 programs with 10 sections each or alternatively switch over to 1 program with 20 sections
- - The time interval of single program sections can be adjusted up to a maximum of 99:59 hours or 999:59 hours. This adjustment applies to all program sections
- Integrated week programm timer with real time function
- Adjustable time functions and programmable temperature ramp function via program editor - Digital temperature setting with an accuracy of one degree
- Adjustable fan speed
- Adjustable front ventilation flap slide and rear exhaust ø 50 mm (1.97 inch)
- Elapsed time indicator
- Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm
- RS 422 interface for communication software APT-COM™ DataControlSystem, or switch over to printer output with RS 232 / RS 422 interface converter
- Units up to 115 liters are stackable
- 2 chrome-plated racks included
- BINDER test certificate
Technical Specifications
Exterior dimensions |
|
Width (mm/inch) |
834 / 32.8 |
Height (inclusive feet) (mm/inch) |
795 / 31.5 |
Depth (mm/inch) |
685 / 27.0 |
Plus door handle (mm/inch) |
50 / 2.0 |
Wall clearance (mm/inch) |
100 / 3.9 |
Wall clearance with open door (mm/inch) |
160 / 6.3 |
Exhaust duct outer- Ø (mm/inch) |
100 / 3.9 |
Steam space volume (l/cu.ft.) |
156 / 5.5 |
Interior dimensions |
|
Width (mm/inch) |
602 / 23.7 |
Height (mm/inch) |
435 / 17.1 |
Depth (mm/inch) |
435 / 17.1 |
Interior volume (l/cu.ft.) |
115 / 4.1 |
Racks, chrome-plated (number standard/max.) |
2 / 5 |
Load per rack (kg/lbs.) |
20 / 44 |
Permitted total load (kg/lbs.) |
50 / 110 |
Weight of the unit (empty) (kg/lbs.) |
90 / 199 |
Temperature data |
|
Temperature range, 5 °C (41 °F) above ambient up to (°C / °F) |
300 / 572 |
Temperature variation |
|
at 70 °C (± °C) |
1.5 |
at 150 °C (± °C) |
2.5 |
at 300 °C (± °C) |
4 |
Temperature fluctuation (± °C) |
0.3 |
Heating-up time 2) |
|
to 70 °C (Min.) |
7 |
to 150 °C (Min.) |
17 |
to 300 °C (Min.) |
44 |
Recov. time after door was opened for 30 sec. 2) |
|
at 70 °C (Min.) |
1 |
at 150 °C (Min.) |
3 |
at 300 °C (Min.) |
6 |
Air change data |
|
Air change (approx. x/min.) |
3 |
Air circulation (approx. x/min.) |
40 |
Exhaust air volume flow (approx. L/Min. m³/h) |
400 (24.0) |
Air flow velocity (m/sec) |
0.8 - 1.2 |
Highest permitted solvent quantity (at T -180 °C, M-100g/mol, U-40g/m3, K=0,5) (g) |
6.65 |
Electrical data |
|
Housing protection acc. to EN 60529 |
IP 33 |
Nominal voltage (±10 %) 50 / 60 Hz (V) |
230 |
Nominal power (kW) |
2.9 |
Energy consumption |
|
at 70 °C (W) |
388 |
at 150 °C (W) |
1098 |
at 300 °C (W) |
2023 |
Contact us on 01939 291200 or by email info@sciquip.co.uk