FDL 115
Safety drying oven with first class safety features

FDL 115

This series presents the perfect environment for solvent-based samples. The interior chamber has a symmetrical airflow with defined flow velocities, free of silicone and dust, safeguarded by a high-efficiency filter cartridge. FDL thus complies with all EN 1539 and ISO 3215 requirements and its program with the intelligent fresh air monitoring feature offers maximum occupational safety.

  • Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
  • All safety features acording to EN 1539
  • Temperature range 5 °C (32 ºF) above ambient temperature up to 300 °C (572 °F)
  • MP Controller, 2 programs with 10 sections each or alternatively switch over to 1 program with 20 sections
  • - The time interval of single program sections can be adjusted up to a maximum of 99:59 hours or 999:59 hours. This adjustment applies to all program sections
    - Integrated week programm timer with real time function
    - Adjustable time functions and programmable temperature ramp function via program editor
  • Digital temperature setting with an accuracy of one degree
  • Adjustable fan speed
  • Adjustable front ventilation flap slide and rear exhaust ø 50 mm (1.97 inch)
  • Elapsed time indicator
  • Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm
  • RS 422 interface for communication software APT-COM™ DataControlSystem, or switch over to printer output with RS 232 / RS 422 interface converter
  • Units up to 115 liters are stackable
  • 2 chrome-plated racks included
  • BINDER test certificate

 

 

 

Technical Specifications

 

Exterior dimensions

 

Width (mm/inch)

834 / 32.8

Height (inclusive feet) (mm/inch)

795 / 31.5

Depth (mm/inch)

685 / 27.0

Plus door handle (mm/inch)

50 / 2.0

Wall clearance (mm/inch)

100 / 3.9

Wall clearance with open door (mm/inch)

160 / 6.3

Exhaust duct outer- Ø (mm/inch)

100 / 3.9

Steam space volume (l/cu.ft.)

156 / 5.5

Interior dimensions

 

Width (mm/inch)

602 / 23.7

Height (mm/inch)

435 / 17.1

Depth (mm/inch)

435 / 17.1

Interior volume (l/cu.ft.)

115 / 4.1

Racks, chrome-plated (number standard/max.)

2 / 5

Load per rack (kg/lbs.)

20 / 44

Permitted total load (kg/lbs.)

50 / 110

Weight of the unit (empty) (kg/lbs.)

90 / 199

Temperature data

 

Temperature range, 5 °C (41 °F) above ambient up to (°C / °F)

300 / 572

Temperature variation

 

at 70 °C (± °C)

1.5

at 150 °C (± °C)

2.5

at 300 °C (± °C)

4

Temperature fluctuation (± °C)

0.3

Heating-up time 2)

 

to 70 °C (Min.)

7

to 150 °C (Min.)

17

to 300 °C (Min.)

44

Recov. time after door was opened for 30 sec. 2)

 

at 70 °C (Min.)

1

at 150 °C (Min.)

3

at 300 °C (Min.)

6

Air change data

 

Air change (approx. x/min.)

3

Air circulation (approx. x/min.)

40

Exhaust air volume flow (approx. L/Min. m³/h)

400 (24.0)

Air flow velocity (m/sec)

0.8 - 1.2

Highest permitted solvent quantity (at T -180 °C, M-100g/mol, U-40g/m3, K=0,5) (g)

6.65

Electrical data

 

Housing protection acc. to EN 60529

IP 33

Nominal voltage (±10 %) 50 / 60 Hz (V)

230

Nominal power (kW)

2.9

Energy consumption

 

at 70 °C (W)

388

at 150 °C (W)

1098

at 300 °C (W)

2023

 

 

 

Contact us on 01939 291200 or by email info@sciquip.co.uk