MDL 115
Safety drying oven with first class safety features and an expanded temperature range
The MDL series operates at temperatures up to 350 ºC at an airflow of 400l/min, ideal conditions for high temperature testing, e.g. in coil coating applications. The preheating chamber with its special Airflow Design permits homogenous baking processes to be performed with maximum occupational safety in extremely short time, safeguarded by electronic monitoring of fresh air. Individual programming options provide maximum flexibility for completing your jobs.
- Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
- Temperature range 5 °C (32 °F) above ambient temperature up to 350 °C (662 °F)
- All safety features acording to EN 1539
- Heat load 9.0 kW
- Door gasket made of of high temperature resistant silicone
- Rear exhaust connector Ø 100 mm (3.94 inch)
- MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
- Features: o User friendly LCD screen o Easy-to-read menu guide o Integrated electronic chart recorder o Variety of options for the graphic display of process parameters o Real time clock
- Replaceable fresh-air filter cartridge, class F6 (EU6-fine particle filter for paricel 1…10 ?m)
- Independent adjustable temperature safety device class 2 (DIN 12880), with visual and audible temperature alarm
- Fresh-air monitoring with audible and visual alarm and automatic deactivation of heating
- RS 422 interface for use with optional GMP/GLP and FDA guideline 21 CFR Part 11 compliant APT-COM™ DataControlSystem software
- 2 chrome-plated racks included
- BINDER test certificate
Technical Specifications
Exterior dimensions |
|
Width (mm/inch) |
834 / 32.8 |
Height (inclusive feet) (mm/inch) |
800 / 31.5 |
Depth (mm/inch) |
685 / 27.0 |
Plus door handle (mm/inch) |
50 / 2.0 |
Wall clearance (mm/inch) |
100 / 3.9 |
Wall clearance with open door (mm/inch) |
160 / 6.3 |
Exhaust duct outer- Ø (mm/inch) |
100 / 3.9 |
Steam space volume (l/cu.ft.) |
156 / 5.5 |
Interior dimensions |
|
Width (mm/inch) |
602 / 23.6 |
Height (mm/inch) |
435 / 17.1 |
Depth (mm/inch) |
435 / 17.1 |
Interior volume (l/cu.ft.) |
115 / 4.1 |
Racks, chrome-plated (number standard/max.) |
2 / 5 |
Load per rack (kg/lbs.) |
20 / 44 |
Permitted total load (kg/lbs.) |
50 / 110 |
Weight of the unit (empty) (kg/lbs.) |
90 / 199 |
Temperature data |
|
Temperature range, 5 °C (41 °F) above ambient up to (°C / °F) |
350 / 662 |
Temperature variation |
|
at 70 °C (± °C) |
2 |
at 150 °C (± °C) |
3.4 |
at 300 °C (± °C) |
7 |
Temperature variation with door flap |
|
at 70 °C (± °C) |
2 |
at 150 °C (± °C) |
3 |
at 300 °C (± °C) |
8 |
Temperature fluctuation (± °C) |
0.5 |
Heating-up time 2) |
|
to 70 °C (Min.) |
3.5 |
to 150 °C (Min.) |
6 |
to 300 °C (Min.) |
10 |
Recov. time after door was opened for 30 sec.2) |
|
at 70 °C (Min.) |
0.5 |
at 150 °C (Min.) |
2 |
at 300 °C (Min.) |
4 |
Recov. time after door was opened for 30 sec.2) |
|
at 70 °C (Min.) |
0.5 |
at 150 °C (Min.) |
1 |
at 300 °C (Min.) |
2 |
Air change data |
|
Air change (approx. x/min.) |
3 |
Air circulation (approx. x/min.) |
40 |
Exhaust air volume flow (approx. L/Min. m³/h) |
400 (24.0) |
Air flow velocity (m/sec) |
0.8 - 1.2 |
Electrical data |
|
Housing protection acc. to EN 60529 |
IP 33 |
Nominal voltage (±10 %) 50 / 60 Hz (V) |
400 3/N |
Nominal power (kW) |
9.0 |
Energy consumption |
|
at 70 °C (W) |
400 |
at 150 °C (W) |
1130 |
at 300 °C (W) |
2083 |